| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 1000.00 | 2.214 | 2.149 | 
| 500.00 | 1.647 | 1.582 | 
| 250.00 | 1.186 | 1.121 | 
| 125.00 | 0.883 | 0.818 | 
| 62.50 | 0.563 | 0.498 | 
| 31.25 | 0.356 | 0.291 | 
| 15.63 | 0.196 | 0.131 | 
| 0.00 | 0.065 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 81-95% | 88% | 
| EDTA plasma(n=5) | 88-96% | 92% | 
| Heparin plasma(n=5) | 95-107% | 101% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 82-94% | 85-96% | 78-90% | 91-99% | 
| EDTA plasma(n=5) | 78-94% | 97-105% | 88-101% | 82-96% | 
| Heparin plasma(n=5) | 89-102% | 79-94% | 83-96% | 85-106% |