| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 4000.00 | 2.045 | 1.961 |
| 2000.00 | 1.616 | 1.532 |
| 1000.00 | 1.157 | 1.073 |
| 500.00 | 0.848 | 0.764 |
| 250.00 | 0.567 | 0.483 |
| 125.00 | 0.357 | 0.273 |
| 62.50 | 0.213 | 0.129 |
| 0.00 | 0.084 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-95% | 89% |
| EDTA plasma(n=5) | 80-92% | 86% |
| Heparin plasma(n=5) | 85-97% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-98% | 87-103% | 89-97% | 85-94% |
| EDTA plasma(n=5) | 95-102% | 83-96% | 97-103% | 85-92% |
| Heparin plasma(n=5) | 87-97% | 86-92% | 89-103% | 89-98% |