| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 2.045 | 1.944 |
| 2500.00 | 1.647 | 1.546 |
| 1250.00 | 1.088 | 0.987 |
| 625.00 | 0.752 | 0.651 |
| 312.50 | 0.593 | 0.492 |
| 156.25 | 0.365 | 0.264 |
| 78.13 | 0.202 | 0.101 |
| 0.00 | 0.101 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-97% | 88% |
| EDTA plasma(n=5) | 96-109% | 102% |
| Heparin plasma(n=5) | 79-93% | 86% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 78-92% | 96-105% | 87-98% | 95-102% |
| EDTA plasma(n=5) | 83-96% | 86-95% | 90-102% | 81-93% |
| Heparin plasma(n=5) | 81-94% | 96-105% | 92-103% | 82-96% |