| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 400.00 | 2.266 | 2.179 |
| 200.00 | 1.536 | 1.449 |
| 100.00 | 1.035 | 0.948 |
| 50.00 | 0.847 | 0.760 |
| 25.00 | 0.468 | 0.381 |
| 12.50 | 0.387 | 0.300 |
| 6.25 | 0.252 | 0.165 |
| 0.00 | 0.087 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 85-97% | 91% |
| EDTA plasma(n=5) | 79-93% | 86% |
| Heparin plasma(n=5) | 82-95% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 96-105% | 87-96% | 90-98% | 92-101% |
| EDTA plasma(n=5) | 84-99% | 81-95% | 98-107% | 96-101% |
| Heparin plasma(n=5) | 88-95% | 98-107% | 90-99% | 88-97% |