| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 800.00 | 1.967 | 1.881 |
| 400.00 | 1.647 | 1.561 |
| 200.00 | 1.183 | 1.097 |
| 100.00 | 0.907 | 0.821 |
| 50.00 | 0.539 | 0.453 |
| 25.00 | 0.356 | 0.270 |
| 12.50 | 0.189 | 0.103 |
| 0.00 | 0.086 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-95% | 87% |
| EDTA plasma(n=5) | 88-102% | 95% |
| Heparin plasma(n=5) | 90-105% | 100% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 83-97% | 82-96% | 86-97% | 90-101% |
| EDTA plasma(n=5) | 91-99% | 85-93% | 82-95% | 93-99% |
| Heparin plasma(n=5) | 82-94% | 92-104% | 82-96% | 95-102% |