| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 60.00 | 2.135 | 2.046 |
| 30.00 | 1.729 | 1.640 |
| 15.00 | 1.027 | 0.938 |
| 7.50 | 0.836 | 0.747 |
| 3.75 | 0.457 | 0.368 |
| 1.88 | 0.377 | 0.288 |
| 0.94 | 0.230 | 0.141 |
| 0.00 | 0.089 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 92-107% | 99% |
| EDTA plasma(n=5) | 78-92% | 85% |
| Heparin plasma(n=5) | 87-99% | 93% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-94% | 87-96% | 82-93% | 89-103% |
| EDTA plasma(n=5) | 98-107% | 93-102% | 80-93% | 91-100% |
| Heparin plasma(n=5) | 90-98% | 85-92% | 83-96% | 93-102% |