| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.132 | 2.047 |
| 1000.00 | 1.621 | 1.536 |
| 500.00 | 1.255 | 1.170 |
| 250.00 | 0.971 | 0.886 |
| 125.00 | 0.617 | 0.532 |
| 62.50 | 0.324 | 0.239 |
| 31.25 | 0.256 | 0.171 |
| 0.00 | 0.085 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 92-106% | 99% |
| EDTA plasma(n=5) | 77-95% | 86% |
| Heparin plasma(n=5) | 88-103% | 95% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-104% | 87-106% | 83-92% | 82-90% |
| EDTA plasma(n=5) | 85-96% | 96-105% | 81-93% | 93-102% |
| Heparin plasma(n=5) | 89-101% | 84-101% | 95-104% | 88-105% |