| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 2.168 | 2.086 |
| 2500.00 | 1.542 | 1.460 |
| 1250.00 | 1.223 | 1.141 |
| 625.00 | 0.812 | 0.730 |
| 312.50 | 0.538 | 0.456 |
| 156.25 | 0.357 | 0.275 |
| 78.13 | 0.238 | 0.156 |
| 0.00 | 0.082 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 78-92% | 85% |
| EDTA plasma(n=5) | 85-99% | 92% |
| Heparin plasma(n=5) | 87-99% | 93% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-104% | 97-106% | 82-91% | 85-96% |
| EDTA plasma(n=5) | 88-103% | 83-94% | 92-105% | 89-97% |
| Heparin plasma(n=5) | 78-92% | 79-93% | 94-106% | 83-97% |