Products

Goat FGF1(Fibroblast Growth Factor 1, Acidic) ELISA Kit
One-click to copy product information
ELK5975
Size: Price:
48 T $412.00
96 T $588.00

Overview

Product name: Goat FGF1(Fibroblast Growth Factor 1, Acidic) ELISA Kit
Reactivity: Goat
Alternative Names: ECGF; AFGF; ECGFA; ECGFB; FGF-Alpha; FGFA; HBGF1; ECGFB; HBGF1; Acidic Fibroblast Growth Factor; heparin-binding growth factor 1; Endothelial Cell Growth Factor, Beta
Assay Type: Sandwich
Sensitivity: 32 pg/mL
Standard: 5000 pg/mL
Detection Range: 78.13-5000 pg/mL
Sample type: Serum, plasma and other biological fluids.
Assay length: 3.5h
Research Area: Cytokine;Tumor immunity;Infection immunity;
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Goat FGF1. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Goat FGF1. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Goat FGF1, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Goat FGF1 in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (pg/mL) OD Corrected OD
5000.00 2.168 2.086
2500.00 1.542 1.460
1250.00 1.223 1.141
625.00 0.812 0.730
312.50 0.538 0.456
156.25 0.357 0.275
78.13 0.238 0.156
0.00 0.082 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant FGF1 and the recovery rates were calculated by comparing the measured value to the expected amount of FGF1 in samples.
Matrix Recovery range Average
serum(n=5) 78-92% 85%
EDTA plasma(n=5) 85-99% 92%
Heparin plasma(n=5) 87-99% 93%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of FGF1 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 89-104% 97-106% 82-91% 85-96%
EDTA plasma(n=5) 88-103% 83-94% 92-105% 89-97%
Heparin plasma(n=5) 78-92% 79-93% 94-106% 83-97%