| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 50.00 | 2.201 | 2.106 |
| 25.00 | 1.579 | 1.484 |
| 12.50 | 1.115 | 1.020 |
| 6.25 | 0.839 | 0.744 |
| 3.13 | 0.535 | 0.440 |
| 1.57 | 0.335 | 0.240 |
| 0.79 | 0.170 | 0.075 |
| 0.00 | 0.095 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-95% | 87% |
| EDTA plasma(n=5) | 86-99% | 92% |
| Heparin plasma(n=5) | 82-95% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 82-95% | 83-96% | 88-95% | 96-102% |
| EDTA plasma(n=5) | 88-102% | 87-102% | 85-96% | 87-101% |
| Heparin plasma(n=5) | 86-96% | 85-97% | 89-101% | 85-97% |