| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 500.00 | 2.126 | 2.037 |
| 250.00 | 1.472 | 1.383 |
| 125.00 | 1.144 | 1.055 |
| 62.50 | 0.897 | 0.808 |
| 31.25 | 0.521 | 0.432 |
| 15.63 | 0.385 | 0.296 |
| 7.82 | 0.216 | 0.127 |
| 0.00 | 0.089 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 95-103% | 99% |
| EDTA plasma(n=5) | 92-102% | 97% |
| Heparin plasma(n=5) | 81-95% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-102% | 87-102% | 85-96% | 87-101% |
| EDTA plasma(n=5) | 78-93% | 96-102% | 95-103% | 89-97% |
| Heparin plasma(n=5) | 85-94% | 87-96% | 82-95% | 93-102% |