| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 2.268 | 2.185 |
| 2500.00 | 1.613 | 1.530 |
| 1250.00 | 1.118 | 1.035 |
| 625.00 | 0.857 | 0.774 |
| 312.50 | 0.582 | 0.499 |
| 156.25 | 0.327 | 0.244 |
| 78.13 | 0.156 | 0.073 |
| 0.00 | 0.083 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 78-90% | 84% |
| EDTA plasma(n=5) | 90-105% | 97% |
| Heparin plasma(n=5) | 95-107% | 101% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 92-107% | 83-99% | 80-96% | 91-101% |
| EDTA plasma(n=5) | 90-97% | 96-102% | 91-103% | 89-97% |
| Heparin plasma(n=5) | 85-94% | 87-103% | 89-97% | 82-90% |