| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 40.00 | 1.967 | 1.879 |
| 20.00 | 1.379 | 1.291 |
| 10.00 | 1.116 | 1.028 |
| 5.00 | 0.834 | 0.746 |
| 2.50 | 0.553 | 0.465 |
| 1.25 | 0.332 | 0.244 |
| 0.63 | 0.302 | 0.214 |
| 0.00 | 0.088 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-99% | 92% |
| EDTA plasma(n=5) | 88-107% | 94% |
| Heparin plasma(n=5) | 97-105% | 101% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-97% | 88-95% | 92-101% | 87-98% |
| EDTA plasma(n=5) | 91-99% | 85-93% | 82-95% | 93-99% |
| Heparin plasma(n=5) | 79-91% | 87-96% | 83-101% | 82-95% |