| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.207 | 2.120 |
| 500.00 | 1.593 | 1.506 |
| 250.00 | 1.088 | 1.001 |
| 125.00 | 0.782 | 0.695 |
| 62.50 | 0.515 | 0.428 |
| 31.25 | 0.549 | 0.462 |
| 15.63 | 0.142 | 0.055 |
| 0.00 | 0.087 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 82-95% | 88% |
| EDTA plasma(n=5) | 95-107% | 101% |
| Heparin plasma(n=5) | 94-107% | 100% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-102% | 92-101% | 88-102% | 87-102% |
| EDTA plasma(n=5) | 79-97% | 86-92% | 81-93% | 89-97% |
| Heparin plasma(n=5) | 90-103% | 97-106% | 82-90% | 92-101% |