| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.075 | 1.989 |
| 5.00 | 1.651 | 1.565 |
| 2.50 | 1.225 | 1.139 |
| 1.25 | 0.872 | 0.786 |
| 0.63 | 0.507 | 0.421 |
| 0.32 | 0.357 | 0.271 |
| 0.16 | 0.161 | 0.075 |
| 0.00 | 0.086 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-95% | 91% |
| EDTA plasma(n=5) | 82-95% | 88% |
| Heparin plasma(n=5) | 94-107% | 100% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-96% | 82-95% | 93-102% | 87-96% |
| EDTA plasma(n=5) | 86-93% | 91-105% | 87-98% | 89-94% |
| Heparin plasma(n=5) | 93-102% | 93-106% | 87-101% | 91-104% |