| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 20.00 | 1.978 | 1.895 |
| 10.00 | 1.566 | 1.483 |
| 5.00 | 1.217 | 1.134 |
| 2.50 | 0.848 | 0.765 |
| 1.25 | 0.587 | 0.504 |
| 0.63 | 0.256 | 0.173 |
| 0.32 | 0.212 | 0.129 |
| 0.00 | 0.083 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-99% | 93% |
| EDTA plasma(n=5) | 78-93% | 85% |
| Heparin plasma(n=5) | 85-97% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-101% | 85-93% | 89-97% | 85-94% |
| EDTA plasma(n=5) | 82-94% | 85-96% | 78-90% | 91-99% |
| Heparin plasma(n=5) | 87-102% | 85-94% | 87-101% | 89-99% |