| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 20.00 | 1.952 | 1.858 |
| 10.00 | 1.597 | 1.503 |
| 5.00 | 1.073 | 0.979 |
| 2.50 | 0.813 | 0.719 |
| 1.25 | 0.576 | 0.482 |
| 0.63 | 0.356 | 0.262 |
| 0.32 | 0.198 | 0.104 |
| 0.00 | 0.094 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 85-97% | 91% |
| EDTA plasma(n=5) | 96-109% | 102% |
| Heparin plasma(n=5) | 81-95% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 91-99% | 85-93% | 82-95% | 93-99% |
| EDTA plasma(n=5) | 88-97% | 95-106% | 83-98% | 93-105% |
| Heparin plasma(n=5) | 88-101% | 90-98% | 92-106% | 85-97% |