| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.052 | 1.969 |
| 1000.00 | 1.507 | 1.424 |
| 500.00 | 1.282 | 1.199 |
| 250.00 | 0.821 | 0.738 |
| 125.00 | 0.585 | 0.502 |
| 62.50 | 0.327 | 0.244 |
| 31.25 | 0.199 | 0.116 |
| 0.00 | 0.083 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 82-98% | 90% |
| EDTA plasma(n=5) | 82-94% | 88% |
| Heparin plasma(n=5) | 80-97% | 96% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-93% | 91-98% | 97-105% | 91-99% |
| EDTA plasma(n=5) | 88-102% | 86-92% | 81-93% | 89-104% |
| Heparin plasma(n=5) | 81-94% | 81-96% | 93-106% | 82-95% |