| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 500.00 | 1.994 | 1.914 |
| 250.00 | 1.507 | 1.427 |
| 125.00 | 1.116 | 1.036 |
| 62.50 | 0.895 | 0.815 |
| 31.25 | 0.468 | 0.388 |
| 15.63 | 0.325 | 0.245 |
| 7.82 | 0.232 | 0.152 |
| 0.00 | 0.080 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-93% | 86% |
| EDTA plasma(n=5) | 78-90% | 84% |
| Heparin plasma(n=5) | 85-97% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-102% | 86-92% | 81-93% | 89-104% |
| EDTA plasma(n=5) | 92-107% | 83-99% | 80-96% | 91-101% |
| Heparin plasma(n=5) | 85-94% | 87-99% | 85-101% | 79-90% |