| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.043 | 1.965 |
| 500.00 | 1.379 | 1.301 |
| 250.00 | 1.116 | 1.038 |
| 125.00 | 0.828 | 0.750 |
| 62.50 | 0.576 | 0.498 |
| 31.25 | 0.332 | 0.254 |
| 15.63 | 0.142 | 0.064 |
| 0.00 | 0.078 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-92% | 86% |
| EDTA plasma(n=5) | 85-99% | 92% |
| Heparin plasma(n=5) | 83-97% | 90% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 93-101% | 86-92% | 81-93% | 79-95% |
| EDTA plasma(n=5) | 98-105% | 92-101% | 85-105% | 87-98% |
| Heparin plasma(n=5) | 85-92% | 79-96% | 89-100% | 85-96% |