48 T | $412.00 |
96 T | $588.00 |
Concentration (µg/mL) | OD | Corrected OD |
---|---|---|
300.00 | 2.401 | 2.317 |
150.00 | 1.613 | 1.529 |
75.00 | 1.196 | 1.112 |
37.50 | 0.866 | 0.782 |
18.75 | 0.509 | 0.425 |
9.38 | 0.324 | 0.240 |
4.69 | 0.186 | 0.102 |
0.00 | 0.084 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 88-106% | 97% |
EDTA plasma(n=5) | 87-99% | 93% |
Heparin plasma(n=5) | 82-97% | 90% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 88-102% | 91-99% | 83-95% | 89-97% |
EDTA plasma(n=5) | 89-97% | 85-94% | 79-96% | 85-99% |
Heparin plasma(n=5) | 95-103% | 87-98% | 85-92% | 79-96% |