| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 1000.00 | 2.601 | 2.522 | 
| 500.00 | 1.781 | 1.702 | 
| 250.00 | 1.135 | 1.056 | 
| 125.00 | 0.873 | 0.794 | 
| 62.50 | 0.509 | 0.430 | 
| 31.25 | 0.386 | 0.307 | 
| 15.63 | 0.235 | 0.156 | 
| 0.00 | 0.079 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 93-109% | 101% | 
| EDTA plasma(n=5) | 92-113% | 103% | 
| Heparin plasma(n=5) | 84-92% | 88% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 83-89% | 90-112% | 93-105% | 87-117% | 
| EDTA plasma(n=5) | 94-118% | 88-112% | 87-106% | 85-106% | 
| Heparin plasma(n=5) | 90-115% | 81-92% | 90-99% | 82-99% | 