| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 500.00 | 2.038 | 1.949 |
| 250.00 | 1.576 | 1.487 |
| 125.00 | 1.107 | 1.018 |
| 62.50 | 0.847 | 0.758 |
| 31.25 | 0.532 | 0.443 |
| 15.63 | 0.379 | 0.290 |
| 7.82 | 0.211 | 0.122 |
| 0.00 | 0.089 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-96% | 88% |
| EDTA plasma(n=5) | 80-92% | 86% |
| Heparin plasma(n=5) | 78-95% | 86% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 97-105% | 86-98% | 84-93% | 89-102% |
| EDTA plasma(n=5) | 85-94% | 83-96% | 87-98% | 87-100% |
| Heparin plasma(n=5) | 88-99% | 81-93% | 89-97% | 85-94% |