| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.052 | 1.948 |
| 500.00 | 1.586 | 1.482 |
| 250.00 | 1.138 | 1.034 |
| 125.00 | 0.951 | 0.847 |
| 62.50 | 0.611 | 0.507 |
| 31.25 | 0.351 | 0.247 |
| 15.63 | 0.225 | 0.121 |
| 0.00 | 0.104 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 78-92% | 85% |
| EDTA plasma(n=5) | 88-106% | 97% |
| Heparin plasma(n=5) | 81-95% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-97% | 91-102% | 81-97% | 93-107% |
| EDTA plasma(n=5) | 85-103% | 82-96% | 87-105% | 90-99% |
| Heparin plasma(n=5) | 88-102% | 86-92% | 81-93% | 79-87% |