| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 20.00 | 1.997 | 1.899 |
| 10.00 | 1.573 | 1.475 |
| 5.00 | 1.085 | 0.987 |
| 2.50 | 0.889 | 0.791 |
| 1.25 | 0.586 | 0.488 |
| 0.63 | 0.334 | 0.236 |
| 0.32 | 0.134 | 0.036 |
| 0.00 | 0.098 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-95% | 89% |
| EDTA plasma(n=5) | 79-93% | 86% |
| Heparin plasma(n=5) | 92-105% | 98% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-97% | 83-96% | 91-103% | 96-105% |
| EDTA plasma(n=5) | 89-105% | 90-101% | 87-98% | 85-92% |
| Heparin plasma(n=5) | 78-92% | 79-93% | 91-98% | 89-102% |