| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 200.00 | 2.138 | 2.035 |
| 100.00 | 1.621 | 1.518 |
| 50.00 | 1.173 | 1.070 |
| 25.00 | 0.873 | 0.770 |
| 12.50 | 0.457 | 0.354 |
| 6.25 | 0.351 | 0.248 |
| 3.13 | 0.157 | 0.054 |
| 0.00 | 0.103 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-99% | 93% |
| EDTA plasma(n=5) | 92-105% | 98% |
| Heparin plasma(n=5) | 83-95% | 89% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-98% | 92-105% | 95-107% | 82-96% |
| EDTA plasma(n=5) | 96-105% | 88-96% | 87-98% | 93-101% |
| Heparin plasma(n=5) | 88-102% | 86-92% | 91-103% | 89-97% |