| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 0.227 | |
| 2500.00 | 0.397 | |
| 1250.00 | 0.611 | |
| 625.00 | 0.857 | |
| 312.50 | 1.119 | |
| 156.25 | 1.511 | |
| 78.13 | 1.789 | |
| 0.00 | 2.123 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-101% | 87% |
| EDTA plasma(n=5) | 86-94% | 90% |
| Heparin plasma(n=5) | 81-93% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-93% | 87-96% | 87-98% | 95-102% |
| EDTA plasma(n=5) | 88-95% | 96-102% | 89-103% | 89-96% |
| Heparin plasma(n=5) | 81-93% | 95-106% | 83-98% | 93-105% |