Human OB(Obestatin) ELISA Kit
One-click to copy product information
ELK5409
Size: Price:
48 T $320.00
96 T $458.00

Overview

Product name: Human OB(Obestatin) ELISA Kit
Reactivity: Human
Alternative Names: Obestatin
Assay Type: Competitive Inhibition
Sensitivity: 25.3 pg/mL
Standard: 5000 pg/mL
Detection Range: 78.13-5000 pg/mL
Sample type: serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay length: 2h
Research Area: Metabolic pathway;Endocrinology;Hepatology;Gastroenterology;Nutrition metabolism;Hormone metabolism;
Test principle: This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with Human OB. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human OB. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human OB in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (pg/mL) OD Corrected OD
5000.00 0.197
2500.00 0.385
1250.00 0.611
625.00 0.823
312.50 1.112
156.25 1.583
78.13 1.796
0.00 2.112

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant OB and the recovery rates were calculated by comparing the measured value to the expected amount of OB in samples.
Matrix Recovery range Average
serum(n=5) 88-103% 95%
EDTA plasma(n=5) 77-95% 86%
Heparin plasma(n=5) 92-106% 95%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of OB and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 85-97% 84-100% 95-102% 88-103%
EDTA plasma(n=5) 86-94% 91-105% 87-99% 95-100%
Heparin plasma(n=5) 89-97% 83-103% 98-105% 92-101%