| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 50.00 | 2.166 | 2.073 |
| 25.00 | 1.677 | 1.584 |
| 12.50 | 1.147 | 1.054 |
| 6.25 | 0.828 | 0.735 |
| 3.13 | 0.576 | 0.483 |
| 1.57 | 0.368 | 0.275 |
| 0.79 | 0.241 | 0.148 |
| 0.00 | 0.093 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 93-107% | 100% |
| EDTA plasma(n=5) | 87-99% | 93% |
| Heparin plasma(n=5) | 85-99% | 92% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-98% | 97-104% | 92-101% | 82-96% |
| EDTA plasma(n=5) | 86-96% | 85-97% | 89-101% | 85-97% |
| Heparin plasma(n=5) | 89-97% | 85-94% | 79-96% | 85-99% |