Products

Rabbit MMP10(Matrix Metalloproteinase 10) ELISA Kit
One-click to copy product information
ELK5377
Size: Price:
48 T $412.00
96 T $588.00

Overview

Product name: Rabbit MMP10(Matrix Metalloproteinase 10) ELISA Kit
Reactivity: Rabbit
Alternative Names: STMY2; SL2; Stromelysin 2; Transin-2
Assay Type: Sandwich
Sensitivity: 25 pg/mL
Standard: 4000 pg/mL
Detection Range: 62.5-4000 pg/mL
Sample type: Serum, plasma, tissue homogenates and other biological fluids
Assay length: 3.5h
Research Area: Tumor immunity;Infection immunity;Cardiovascular biology;
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Rabbit MMP10. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Rabbit MMP10. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Rabbit MMP10, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Rabbit MMP10 in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (pg/mL) OD Corrected OD
4000.00 2.128 2.033
2000.00 1.700 1.605
1000.00 1.227 1.132
500.00 0.928 0.833
250.00 0.553 0.458
125.00 0.318 0.223
62.50 0.216 0.121
0.00 0.095 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant MMP10 and the recovery rates were calculated by comparing the measured value to the expected amount of MMP10 in samples.
Matrix Recovery range Average
serum(n=5) 79-93% 86%
EDTA plasma(n=5) 92-105% 98%
Heparin plasma(n=5) 83-95% 89%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of MMP10 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 87-95% 85-98% 82-97% 92-101%
EDTA plasma(n=5) 85-102% 84-99% 97-101% 89-102%
Heparin plasma(n=5) 86-93% 91-103% 87-98% 92-99%