| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 4000.00 | 2.128 | 2.033 |
| 2000.00 | 1.700 | 1.605 |
| 1000.00 | 1.227 | 1.132 |
| 500.00 | 0.928 | 0.833 |
| 250.00 | 0.553 | 0.458 |
| 125.00 | 0.318 | 0.223 |
| 62.50 | 0.216 | 0.121 |
| 0.00 | 0.095 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 79-93% | 86% |
| EDTA plasma(n=5) | 92-105% | 98% |
| Heparin plasma(n=5) | 83-95% | 89% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-95% | 85-98% | 82-97% | 92-101% |
| EDTA plasma(n=5) | 85-102% | 84-99% | 97-101% | 89-102% |
| Heparin plasma(n=5) | 86-93% | 91-103% | 87-98% | 92-99% |