| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.389 | 2.295 |
| 500.00 | 1.507 | 1.413 |
| 250.00 | 0.993 | 0.899 |
| 125.00 | 0.671 | 0.577 |
| 62.50 | 0.538 | 0.444 |
| 31.25 | 0.324 | 0.230 |
| 15.63 | 0.217 | 0.123 |
| 0.00 | 0.094 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-95% | 87% |
| EDTA plasma(n=5) | 92-105% | 98% |
| Heparin plasma(n=5) | 84-96% | 90% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 81-94% | 96-105% | 92-103% | 82-96% |
| EDTA plasma(n=5) | 85-97% | 86-102% | 87-99% | 99-105% |
| Heparin plasma(n=5) | 95-104% | 82-94% | 86-96% | 81-97% |