| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 500.00 | 1.898 | 1.817 |
| 250.00 | 1.527 | 1.446 |
| 125.00 | 0.996 | 0.915 |
| 62.50 | 0.727 | 0.646 |
| 31.25 | 0.468 | 0.387 |
| 15.63 | 0.315 | 0.234 |
| 7.82 | 0.236 | 0.155 |
| 0.00 | 0.081 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-97% | 88% |
| EDTA plasma(n=5) | 95-108% | 101% |
| Heparin plasma(n=5) | 83-97% | 90% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 93-102% | 81-96% | 93-106% | 82-95% |
| EDTA plasma(n=5) | 85-96% | 87-101% | 89-102% | 85-98% |
| Heparin plasma(n=5) | 83-103% | 82-93% | 86-95% | 81-96% |