| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 1.990 | 1.907 |
| 5.00 | 1.602 | 1.519 |
| 2.50 | 1.174 | 1.091 |
| 1.25 | 0.883 | 0.800 |
| 0.63 | 0.543 | 0.460 |
| 0.32 | 0.276 | 0.193 |
| 0.16 | 0.227 | 0.144 |
| 0.00 | 0.083 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 82-90% | 86% |
| EDTA plasma(n=5) | 84-92% | 87% |
| Heparin plasma(n=5) | 95-104% | 98% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-95% | 91-106% | 87-100% | 95-101% |
| EDTA plasma(n=5) | 87-98% | 92-105% | 95-107% | 82-96% |
| Heparin plasma(n=5) | 95-104% | 82-94% | 86-96% | 81-97% |