| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 2.266 | 2.171 |
| 2500.00 | 1.647 | 1.552 |
| 1250.00 | 1.069 | 0.974 |
| 625.00 | 0.826 | 0.731 |
| 312.50 | 0.433 | 0.338 |
| 156.25 | 0.273 | 0.178 |
| 78.13 | 0.211 | 0.116 |
| 0.00 | 0.095 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-105% | 97% |
| EDTA plasma(n=5) | 88-102% | 95% |
| Heparin plasma(n=5) | 80-92% | 86% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-104% | 84-98% | 95-105% | 88-102% |
| EDTA plasma(n=5) | 85-96% | 87-101% | 89-102% | 85-98% |
| Heparin plasma(n=5) | 93-105% | 82-101% | 91-103% | 95-102% |