| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 100.00 | 2.532 | 2.446 |
| 50.00 | 1.731 | 1.645 |
| 25.00 | 1.132 | 1.046 |
| 12.50 | 0.868 | 0.782 |
| 6.25 | 0.634 | 0.548 |
| 3.13 | 0.340 | 0.254 |
| 1.57 | 0.142 | 0.056 |
| 0.00 | 0.086 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-95% | 87% |
| EDTA plasma(n=5) | 83-97% | 90% |
| Heparin plasma(n=5) | 78-92% | 85% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-94% | 87-101% | 89-99% | 85-95% |
| EDTA plasma(n=5) | 86-93% | 91-101% | 87-98% | 95-104% |
| Heparin plasma(n=5) | 82-94% | 85-96% | 78-90% | 91-99% |