| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 2.157 | 2.074 |
| 2500.00 | 1.569 | 1.486 |
| 1250.00 | 1.285 | 1.202 |
| 625.00 | 0.989 | 0.906 |
| 312.50 | 0.580 | 0.497 |
| 156.25 | 0.312 | 0.229 |
| 78.13 | 0.248 | 0.165 |
| 0.00 | 0.083 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 88-102% | 95% |
| EDTA plasma(n=5) | 78-92% | 85% |
| Heparin plasma(n=5) | 80-97% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 95-102% | 85-96% | 90-98% | 82-93% |
| EDTA plasma(n=5) | 89-103% | 79-92% | 88-97% | 85-97% |
| Heparin plasma(n=5) | 95-107% | 92-103% | 81-99% | 90-98% |