| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.532 | 2.455 |
| 5.00 | 1.828 | 1.751 |
| 2.50 | 1.106 | 1.029 |
| 1.25 | 0.596 | 0.519 |
| 0.63 | 0.395 | 0.318 |
| 0.32 | 0.305 | 0.228 |
| 0.16 | 0.179 | 0.102 |
| 0.00 | 0.077 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 82-95% | 88% |
| EDTA plasma(n=5) | 80-93% | 86% |
| Heparin plasma(n=5) | 95-108% | 101% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-100% | 79-91% | 89-103% | 82-95% |
| EDTA plasma(n=5) | 93-101% | 97-103% | 91-99% | 90-98% |
| Heparin plasma(n=5) | 92-107% | 83-99% | 80-96% | 91-101% |