| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 1.961 | 1.858 |
| 2500.00 | 1.594 | 1.491 |
| 1250.00 | 1.143 | 1.040 |
| 625.00 | 0.971 | 0.868 |
| 312.50 | 0.563 | 0.460 |
| 156.25 | 0.300 | 0.197 |
| 78.13 | 0.226 | 0.123 |
| 0.00 | 0.103 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 85-97% | 91% |
| EDTA plasma(n=5) | 91-105% | 98% |
| Heparin plasma(n=5) | 81-95% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-97% | 85-101% | 84-98% | 85-93% |
| EDTA plasma(n=5) | 88-97% | 86-97% | 85-103% | 89-97% |
| Heparin plasma(n=5) | 79-96% | 85-99% | 79-93% | 89-97% |