| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 4000.00 | 1.946 | 1.842 |
| 2000.00 | 1.517 | 1.413 |
| 1000.00 | 1.185 | 1.081 |
| 500.00 | 0.895 | 0.791 |
| 250.00 | 0.587 | 0.483 |
| 125.00 | 0.361 | 0.257 |
| 62.50 | 0.236 | 0.132 |
| 0.00 | 0.104 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-93% | 90% |
| EDTA plasma(n=5) | 80-90% | 85% |
| Heparin plasma(n=5) | 84-106% | 100% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 83-97% | 86-97% | 90-98% | 80-93% |
| EDTA plasma(n=5) | 82-94% | 85-96% | 78-90% | 91-99% |
| Heparin plasma(n=5) | 82-96% | 85-98% | 95-104% | 82-96% |