| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 1.941 | 1.846 |
| 1000.00 | 1.548 | 1.453 |
| 500.00 | 1.250 | 1.155 |
| 250.00 | 0.826 | 0.731 |
| 125.00 | 0.575 | 0.480 |
| 62.50 | 0.299 | 0.204 |
| 31.25 | 0.211 | 0.116 |
| 0.00 | 0.095 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 94-104% | 99% |
| EDTA plasma(n=5) | 81-115% | 98% |
| Heparin plasma(n=5) | 92-115% | 104% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 91-116% | 82-120% | 91-102% | 89-107% |
| EDTA plasma(n=5) | 87-119% | 80-90% | 92-114% | 93-105% |
| Heparin plasma(n=5) | 83-100% | 89-103% | 88-108% | 91-104% |