| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 1.991 | 1.923 |
| 5.00 | 1.598 | 1.530 |
| 2.50 | 1.133 | 1.065 |
| 1.25 | 0.804 | 0.736 |
| 0.63 | 0.524 | 0.456 |
| 0.31 | 0.330 | 0.262 |
| 0.16 | 0.229 | 0.161 |
| 0.00 | 0.068 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-100% | 95% |
| EDTA plasma(n=5) | 86-113% | 100% |
| Heparin plasma(n=5) | 92-110% | 101% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 82-94% | 85-98% | 94-119% | 91-99% |
| EDTA plasma(n=5) | 85-96% | 91-98% | 86-96% | 88-115% |
| Heparin plasma(n=5) | 94-104% | 86-98% | 91-109% | 92-117% |