48 T | $412.00 |
96 T | $588.00 |
Concentration (mIU/mL) | OD | Corrected OD |
---|---|---|
100.00 | 1.842 | 1.755 |
50.00 | 1.545 | 1.458 |
25.00 | 1.053 | 0.966 |
12.50 | 0.871 | 0.784 |
6.25 | 0.655 | 0.568 |
3.13 | 0.377 | 0.290 |
1.57 | 0.213 | 0.126 |
0.00 | 0.087 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 83-99% | 95% |
EDTA plasma(n=5) | 80-96% | 88% |
Heparin plasma(n=5) | 87-99% | 91% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 86-101% | 86-101% | 80-101% | 87-96% |
EDTA plasma(n=5) | 79-96% | 85-99% | 79-93% | 89-97% |
Heparin plasma(n=5) | 93-105% | 96-105% | 87-98% | 95-102% |