| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 20.00 | 1.952 | 1.858 |
| 10.00 | 1.550 | 1.456 |
| 5.00 | 0.989 | 0.895 |
| 2.50 | 0.766 | 0.672 |
| 1.25 | 0.522 | 0.428 |
| 0.63 | 0.238 | 0.144 |
| 0.31 | 0.213 | 0.119 |
| 0.00 | 0.094 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 86-94% | 90% |
| EDTA plasma(n=5) | 87-99% | 93% |
| Heparin plasma(n=5) | 93-108% | 101% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 83-105% | 92-115% | 83-117% | 93-110% |
| EDTA plasma(n=5) | 91-98% | 84-93% | 85-91% | 94-100% |
| Heparin plasma(n=5) | 94-101% | 84-93% | 86-116% | 95-112% |