| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 2000.00 | 1.988 | 1.892 | 
| 1000.00 | 1.647 | 1.551 | 
| 500.00 | 1.177 | 1.081 | 
| 250.00 | 0.772 | 0.676 | 
| 125.00 | 0.469 | 0.373 | 
| 62.50 | 0.332 | 0.236 | 
| 31.25 | 0.253 | 0.157 | 
| 0.00 | 0.096 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 85-108% | 97% | 
| EDTA plasma(n=5) | 86-92% | 89% | 
| Heparin plasma(n=5) | 87-103% | 95% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 91-117% | 87-118% | 87-113% | 95-102% | 
| EDTA plasma(n=5) | 82-92% | 91-102% | 85-102% | 88-117% | 
| Heparin plasma(n=5) | 80-87% | 87-102% | 85-108% | 92-108% | 