48 T | $320.00 |
96 T | $458.00 |
Concentration (pg/mL) | OD | Corrected OD |
---|---|---|
2000.00 | 1.994 | 1.901 |
1000.00 | 1.562 | 1.469 |
500.00 | 1.032 | 0.939 |
250.00 | 0.782 | 0.689 |
125.00 | 0.457 | 0.364 |
62.50 | 0.331 | 0.238 |
31.25 | 0.156 | 0.063 |
0.00 | 0.093 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 82-95% | 88% |
EDTA plasma(n=5) | 95-107% | 101% |
Heparin plasma(n=5) | 90-105% | 97% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 86-93% | 93-101% | 85-98% | 90-101% |
EDTA plasma(n=5) | 91-99% | 83-97% | 87-105% | 87-99% |
Heparin plasma(n=5) | 85-100% | 84-99% | 85-97% | 88-102% |