| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 1.952 | 1.869 |
| 1000.00 | 1.545 | 1.462 |
| 500.00 | 1.088 | 1.005 |
| 250.00 | 0.752 | 0.669 |
| 125.00 | 0.545 | 0.462 |
| 62.50 | 0.256 | 0.173 |
| 31.25 | 0.207 | 0.124 |
| 0.00 | 0.083 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 81-93% | 87% |
| EDTA plasma(n=5) | 83-101% | 87% |
| Heparin plasma(n=5) | 86-94% | 90% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 96-103% | 83-96% | 87-98% | 85-97% |
| EDTA plasma(n=5) | 78-93% | 96-102% | 95-103% | 89-97% |
| Heparin plasma(n=5) | 85-94% | 87-96% | 82-95% | 93-102% |