| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 50.00 | 2.311 | 2.229 |
| 25.00 | 1.655 | 1.573 |
| 12.50 | 1.146 | 1.064 |
| 6.25 | 0.907 | 0.825 |
| 3.13 | 0.529 | 0.447 |
| 1.57 | 0.239 | 0.157 |
| 0.79 | 0.175 | 0.093 |
| 0.00 | 0.082 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-97% | 88% |
| EDTA plasma(n=5) | 82-95% | 88% |
| Heparin plasma(n=5) | 88-99% | 93% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-95% | 98-107% | 90-99% | 88-97% |
| EDTA plasma(n=5) | 92-106% | 93-105% | 85-99% | 90-99% |
| Heparin plasma(n=5) | 88-104% | 84-98% | 95-105% | 88-102% |