| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.160 | 2.058 |
| 5.00 | 1.589 | 1.487 |
| 2.50 | 1.113 | 1.011 |
| 1.25 | 0.751 | 0.649 |
| 0.63 | 0.529 | 0.427 |
| 0.31 | 0.353 | 0.251 |
| 0.16 | 0.226 | 0.124 |
| 0.00 | 0.102 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 89-97% | 93% |
| EDTA plasma(n=5) | 80-115% | 98% |
| Heparin plasma(n=5) | 94-114% | 104% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 84-111% | 90-110% | 93-99% | 83-120% |
| EDTA plasma(n=5) | 82-90% | 83-116% | 93-107% | 81-112% |
| Heparin plasma(n=5) | 82-111% | 80-116% | 81-113% | 85-93% |