| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.145 | 2.056 |
| 5.00 | 1.519 | 1.430 |
| 2.50 | 1.094 | 1.005 |
| 1.25 | 0.902 | 0.813 |
| 0.63 | 0.383 | 0.294 |
| 0.31 | 0.328 | 0.239 |
| 0.16 | 0.208 | 0.119 |
| 0.00 | 0.089 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-99% | 95% |
| EDTA plasma(n=5) | 83-116% | 100% |
| Heparin plasma(n=5) | 90-111% | 101% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-93% | 89-118% | 89-100% | 90-97% |
| EDTA plasma(n=5) | 95-114% | 88-95% | 87-97% | 94-119% |
| Heparin plasma(n=5) | 84-111% | 80-108% | 92-104% | 89-114% |