| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 2.029 | 1.925 | 
| 5.00 | 1.628 | 1.524 | 
| 2.50 | 1.253 | 1.149 | 
| 1.25 | 0.752 | 0.648 | 
| 0.63 | 0.562 | 0.458 | 
| 0.32 | 0.324 | 0.220 | 
| 0.16 | 0.168 | 0.064 | 
| 0.00 | 0.104 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 83-95% | 89% | 
| EDTA plasma(n=5) | 79-93% | 86% | 
| Heparin plasma(n=5) | 92-105% | 98% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 85-97% | 84-100% | 95-103% | 88-104% | 
| EDTA plasma(n=5) | 86-95% | 91-106% | 87-100% | 95-101% | 
| Heparin plasma(n=5) | 95-102% | 83-97% | 88-96% | 96-103% |