| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.127 | 2.039 |
| 5.00 | 1.616 | 1.528 |
| 2.50 | 1.121 | 1.033 |
| 1.25 | 0.906 | 0.818 |
| 0.63 | 0.551 | 0.463 |
| 0.31 | 0.394 | 0.306 |
| 0.16 | 0.201 | 0.113 |
| 0.00 | 0.088 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-114% | 99% |
| EDTA plasma(n=5) | 83-126% | 105% |
| Heparin plasma(n=5) | 94-118% | 106% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-95% | 87-100% | 92-109% | 91-109% |
| EDTA plasma(n=5) | 93-117% | 83-103% | 81-88% | 83-113% |
| Heparin plasma(n=5) | 83-92% | 80-105% | 82-96% | 94-102% |